Peak Hop Definition ------------------------ This section describes how to define a peak hopping sequence. A peak hop sequence, ``HOPS`` , is a list of peak hops (i.e magnet moves). :: HOPS=[('Ar40:H1:10, Ar39:AX, Ar36:CDD', 5, 1), ('Ar37:CDD', 5, 1)] A ``peak hop`` is a list of ``iso:detector`` pairs plus two configuration values, #. the number of counts at the current position #. the settling time (s) after magnet positioning and before measurement starts. The first ``iso:detector`` pair of each hop is used for positioning. To specify a non-nominal deflection value for a detector use ``iso:detector:deflection`` :: Ar40:H1:30 If no value is specified and the deflection value had been changed by a previous cycle then the deflection is set to the value stored in the spectrometer configuration file. The following sequence, :: ('Ar40:H1, Ar39:AX, Ar36:CDD', 10, 1) ('Ar40:L2, Ar39:CDD', 20, 5) translates to #. position Ar40 on detector H1, wait 1s and record 10 H1,AX,and CDD measurements. #. After 10 measurements position Ar40 on detector L2, wait 5s, then record 20 measurements.